To calibrate the capacitance and inductance tester to ensure measurement accuracy, the following steps can be followed:

① Connect calibration standards: First, connect the calibration standards (such as components with known capacitance or inductance values) to the testing port of the tester.

② Set testing frequency: Based on the characteristics and calibration requirements of the tested components, set the testing frequency to the calibration frequency, usually 1kHz or 10kHz.

③ Perform calibration operation: Follow the instructions of the tester to perform the calibration operation. This may include adjusting the calibration capacitance value of the tester, calibrating the circuit, etc. Ensure calibration is carried out in the correct steps and sequence.

④ Verify calibration results: After calibration is completed, verification is required. Use calibration standards for retesting and compare the results with those before calibration. Confirm whether the test results meet the requirements of calibration standards, that is, whether the measured values are consistent with the actual values or whether the error is within the allowable range.

⑤ Record calibration results: Record the calibration results, including calibration date, calibration personnel, calibration standards, calibration results, and other information. This helps to track the performance and accuracy of the tester and provide reference when needed.

In addition, to ensure the measurement accuracy of the capacitance and inductance tester, the following points need to be noted:

① Choose an appropriate calibration standard: The calibration standard should have known and accurate capacitance or inductance values, and should match the measurement range of the tester.

② Check the testing environment: Interference from the testing environment can have an impact on the test results. Therefore, before conducting calibration, it is necessary to check whether the testing environment meets the requirements. For example, it is advisable to avoid testing near strong magnetic or electric fields as much as possible to avoid bias in test results.

③ Regular calibration: To ensure the long-term stability and accuracy of the testing instrument, it should be calibrated regularly. The calibration cycle can be determined based on the frequency of use and measurement requirements of the tester.

④ Correct use and maintenance: When using the tester, the correct operating steps and maintenance requirements should be followed. Avoid overloading, overheating, or improper use to extend the lifespan of the tester and maintain its accuracy.

The above steps and precautions can help you calibrate the capacitance and inductance tester and ensure its measurement accuracy. If necessary, it is recommended to refer to the instructions of the tester or contact professional personnel for calibration operations.

ZC-800C正.jpg

ZC-800C Capacitance and inductance tester instrument adopts high-speed microprocessor to synchronously collect the voltage signal and current signal of the tested object, and automatically calculate the capacitance value and inductance value. The on-site measurement capacitor does not need to remove the connecting wire, which simplifies the test process and effectively improves the work efficiency. After the test, the capacitor capacity and other parameters of each phase are automatically calculated, and it is very easy to identify the quality change of the capacitor and the fault of the connecting conductor between devices.

Kvtester Electronics Technology Co.,Ltd. is a high-tech enterprise specializing in power testing, testing, research and development, production, and sales of testing equipment. It has been engaged in the electrical testing industry for many years, and its products are of high quality. We welcome customers to come and purchase. Service hotline: 0086-27-81778799, to learn more, visit the official website: www.kvtester.com